New Functions of the new one compares to the old version temperature data recorders:
-
Auto PDF Report: The new-added function allows the data loggerto generate a PDF report automatically.
-
Password Protection: Optional password protection will keep your datasafe and sound,
-
Double the Battery Life: with a battery life up to 12 months (15℃~23℃/45%RH~75%RH), which doubles battery lifeof the old temperature loggers.
|

|

|

|
RC-5
|
RC-5+
|
RC-5+ with probe
|
Record capacity
|
32,000 Points
|
32,000 Points
|
32,000 Points
|
Temperature Range
|
-30°C ~ +70°C
|
-30°C ~ + 70°C
|
-40°C ~ + 85°C
|
Battery Life
|
6 Month
|
12 Month
|
12 Month
|
Connection Type
|
USB
|
USB
|
USB
|
Auto PDF Report
|
No
|
Yes
|
Yes
|
Restart by press button
|
No
|
Yes
|
Yes
|
Need Software
|
Yes
|
No
|
No
|
Password protection
|
No
|
Yes
|
Yes
|
External sensor
|
No
|
No
|
Yes
|
Package included:
1* ThermElc RC-5+ temperate data logger
1* External temperature probe with cap
1* CR2032 battery in the logger
1* User manual
1* 2-Year calibration certificate

Temperature Data Logger Applied in COVID-19 Vaccine Transportation and Handling
As we know, COVID-19 pandemic outbreak at the end of Year 2019 and spreading all over the world. Billions of people are suffering from the pandemic. Thus equitable access to safe and effective vaccines is critical to ending the COVID-19 pandemic, so it is hugely encouraging to see so many vaccines proving and going into development. WHO is working tirelessly with partners to develop, manufacture and deploy safe and effective vaccines.
How to carefully handling the vaccines and monitoring the temperature of the vaccines is very crucial during the whole procedure of the transportation and storage.
Vaccine Cold Chain
A cold chain is a temperature-controlled supply chain that includes all vaccine-related equipment and procedures. The cold chain begins with the cold storage unit at the manufacturing plant, extends to the transport and delivery of the vaccine and proper storage at the provider facility, and ends with administration of the vaccine to the patient. Manufacturers, distributors, public health staff, and health care providers share responsibility to ensure the vaccine cold chain is maintained from the time vaccines are manufactured until they are administered.
When transporting COVID-19 vaccine, validate the temperature whenever the storage container is opened. Minimize the total transport time to reduce potential risk for a temperature excursion due to a storage unit or thermal packaging system failure. A temperature monitoring and data recorder must be utilized in transport.
Our RC-5+ temperature data logger with probe for vaccines plays a very vital role in the process. Its external probe can be inserted in the ice-box of the vaccine without open the cover to avoid temperature sudden change in the cold chain transportation. That helps the validity of the vaccine and ensure its safety to the patients.

Instruction for Initial Use of Temperature Logger with Probe
-
Download the free software of the loggers(Mac or Windows OS version).
-
Connect the data loggerto your PC.
-
Open the software and set the parameters as you wish.
- Save the parameter and disconnect the digital data logger.
- Press start button to enable the data recorder.
Meanings of the Setting Parameters:
Log Clock: Not configurable. But the time will sync with that on the computer after our user clicks the Save Parameter button under the Parameter tab.
S/N: Not configurable.
Trip Description: Configurable. The info entered here will be shown on the report, it can be used to differentiate one logger from another.
Log interval: Configurable. This means how often our user wants the logger to perform a reading, it can be 10S, 1minute, or 1day.
Start Delay(H): Configurable. This means after you press the start button on the logger, how long it would wait to officially start logging.
Press button: Configurable. This means after we press the start button for 4 seconds on the logger to start logging. We can still press this button to restart the logger.
Temperature Unit: Celsius or Fahrenheit Optional.
High/low limit: Configurable. The upper and lower alarming point you want to set.
Specifications
Temperature Measurement Range
|
Internal sensor: -30°C ~+70°C(-22℉~158℉)
External probe: -40°C ~+85°C(-40°F to 185°F)
|
Accuracy
|
±0.5℃/±0.9℉(-20℃ ~ 40℃)
±1℃/±1.8℉(others)
|
Resolution
|
0.1℃/℉
|
Data Capacity
|
32,000 readings
|
Logging Interval
|
10sec - 24hr
|
Communication Interface
|
USB
|
Report Format
|
Auto-generated PDF report
Software: PDF/XLS/TXT
|
Software
|
ElitechLog (Windows&macOS)
|
Power Supply
|
Single-use CR2032 replaceable battery
|
Shelf Life
|
1 Year
|
Protection Grade
|
IP67
|
Certification
|
EN12830, CE, RoHS
|
Dimensions
|
80 x 33.5 x 14 mm
|
Weight
|
20g
|